LIVE · cybersecurity feed
Live wire
cve recordhigh

CVE-2019-25722

Published
CVSS7.6
Severityhigh
WeaknessCWE-798
ExploitedNot in CISA KEV

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

Description

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.